T1 - A comparison of hemoglobin measured by co-oximetry and central laboratory during major spine fusion surgery
AU - Carabini, Louanne M.
AU - Navarre, William J.
AU - Ault, Michael L.
AU - Bebawy, John F.
AU - Gupta, Dhanesh K.
N1 - Publisher Copyright:
© 2014 International Anesthesia Research Society.
PY - 2015
Y1 - 2015
N2 - BACKGROUND: Many factors affect the accuracy of hemoglobin concentration values. In this study, we evaluated whether the hemoglobin concentration obtained by means of arterial blood gas (ABG) co-oximetry and complete blood count (CBC) central laboratory techniques clinically correlate when using simultaneous measurements of hemoglobin concentration obtained during complex spine fusion surgery. METHODS: Three hundred forty-eight patients who underwent spinal fusion of >3 bony levels between September 2006 and September 2010, with concurrent ABG and CBC samples, were identified. The mean difference between pairs of measured hemoglobin values was determined using limits of agreement analysis. Error grid analysis was used to delineate correlation of samples in relation to hemoglobin values within the range considered for transfusion. RESULTS: The median difference (ABG-CBC) between the measured hemoglobin values was 0.4 g/dL (95% confidence interval [CI], 0.35-0.40 g/dL; P < 0.0001). Limits of agreement analysis correcting for repeated observations in multiple patients demonstrated that the mean difference between measured hemoglobin values (i.e., bias) was 0.4 g/dL (95% CI, 0.36-0.41 g/dL), and the 95% limits of agreement of the difference between paired measurements were -0.70 to 1.47 g/dL. The magnitude of the difference between the measured hemoglobin values was >0.5 g/dL in 44.5% of patients (95% CI, 42.2%-46.8%); however, 6.8% (95% CI, 5.8%- 8.1%) of paired measurements had a difference of >1.0 g/dL. There was only fair-to-moderate agreement between the CBC and ABG values within the clinically significant range of hemoglobin values of 7 to 10 g/dL (Cohen κ = 0.39; 95% CI, 0.33-0.45). CONCLUSIONS: The hemoglobin values obtained from ABG and CBC cannot be used interchangeably when verifying accuracy of novel point-of-care hemoglobin measurement modalities or when managing a patient with critical blood loss.
AB - BACKGROUND: Many factors affect the accuracy of hemoglobin concentration values. In this study, we evaluated whether the hemoglobin concentration obtained by means of arterial blood gas (ABG) co-oximetry and complete blood count (CBC) central laboratory techniques clinically correlate when using simultaneous measurements of hemoglobin concentration obtained during complex spine fusion surgery. METHODS: Three hundred forty-eight patients who underwent spinal fusion of >3 bony levels between September 2006 and September 2010, with concurrent ABG and CBC samples, were identified. The mean difference between pairs of measured hemoglobin values was determined using limits of agreement analysis. Error grid analysis was used to delineate correlation of samples in relation to hemoglobin values within the range considered for transfusion. RESULTS: The median difference (ABG-CBC) between the measured hemoglobin values was 0.4 g/dL (95% confidence interval [CI], 0.35-0.40 g/dL; P < 0.0001). Limits of agreement analysis correcting for repeated observations in multiple patients demonstrated that the mean difference between measured hemoglobin values (i.e., bias) was 0.4 g/dL (95% CI, 0.36-0.41 g/dL), and the 95% limits of agreement of the difference between paired measurements were -0.70 to 1.47 g/dL. The magnitude of the difference between the measured hemoglobin values was >0.5 g/dL in 44.5% of patients (95% CI, 42.2%-46.8%); however, 6.8% (95% CI, 5.8%- 8.1%) of paired measurements had a difference of >1.0 g/dL. There was only fair-to-moderate agreement between the CBC and ABG values within the clinically significant range of hemoglobin values of 7 to 10 g/dL (Cohen κ = 0.39; 95% CI, 0.33-0.45). CONCLUSIONS: The hemoglobin values obtained from ABG and CBC cannot be used interchangeably when verifying accuracy of novel point-of-care hemoglobin measurement modalities or when managing a patient with critical blood loss.
UR - http://www.scopus.com/inward/record.url?scp=84923919144&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84923919144&partnerID=8YFLogxK
U2 - 10.1213/ANE.0000000000000418
DO - 10.1213/ANE.0000000000000418
M3 - Article
C2 - 25185592
AN - SCOPUS:84923919144
SN - 0003-2999
VL - 120
SP - 60
EP - 65
JO - Anesthesia and analgesia
JF - Anesthesia and analgesia
IS - 1
ER -