T1 - A thin-film multichannel electrode for muscle recording and stimulation in neuroprosthetics applications
AU - Muceli, Silvia
AU - Poppendieck, Wigand
AU - Hoffmann, Klaus Peter
AU - Dosen, Strahinja
AU - Benito-León, Julián
AU - Barroso, Filipe Oliveira
AU - Pons, Jose L
AU - Farina, Dario
N1 - Funding Information:
This study was financed by the European Projects EXTEND (H2020-ICT-23-2017-779982) and NeuroTREMOR (FP7-ICT-2011-7-287739). The funding agency was not involved in the collection, analyses, and interpretation of data or writing and publication of this article.
Publisher Copyright:
© 2019 IOP Publishing Ltd.
PY - 2019
Y1 - 2019
N2 - Objective. We propose, design and test a novel thin-film multichannel electrode that can be used for both recording from and stimulating a muscle in acute implants. Approach. The system is built on a substrate of polyimide and contains 12 recording and three stimulation sites made of platinum. The structure is 420 μm wide, 20 μm thick and embeds the recording and stimulation contacts on the two sides of the polyimide over an approximate length of 2 cm. We show representative applications in healthy individuals as well as tremor patients. The designed system was tested by a psychometric characterization of the stimulation contacts in six tremor patients and three healthy individuals determining the perception threshold and current limit as well as the success rate in discriminating elicited sensations (electrotactile feedback). Also, we investigated the possibility of using the intramuscular electrode for reducing tremor in one patient by electrical stimulation delivered with timing based on the electromyographic activity recorded with the same electrode. Main results. In the tremor patients, the current corresponding to the perception threshold and the current limit were 0.7 ± 0.2 and 1.4 ± 0.7 mA for the wrist flexor muscles and 0.4 ± 0.2 and 1.5 ± 0.7 mA for the extensors. In one patient, closed-loop stimulation resulted in a decrease of the tremor power >50%. In healthy individuals the perception threshold and current limits were 0.9 ± 0.6 and 2.1 ± 0.6 mA for the extensor carpi radialis muscle. The subjects could distinguish four or six stimulation patterns (two or three stimulation sites × two stimulation current amplitudes) with true positive rate >80% (two subjects) and >60% (one subject), respectively. Significance. The proposed electrode provides a compact multichannel interface for recording electromyogram and delivering electrical stimulation in applications such as neuroprostheses for tremor suppression and closed-loop myoelectric prostheses.
AB - Objective. We propose, design and test a novel thin-film multichannel electrode that can be used for both recording from and stimulating a muscle in acute implants. Approach. The system is built on a substrate of polyimide and contains 12 recording and three stimulation sites made of platinum. The structure is 420 μm wide, 20 μm thick and embeds the recording and stimulation contacts on the two sides of the polyimide over an approximate length of 2 cm. We show representative applications in healthy individuals as well as tremor patients. The designed system was tested by a psychometric characterization of the stimulation contacts in six tremor patients and three healthy individuals determining the perception threshold and current limit as well as the success rate in discriminating elicited sensations (electrotactile feedback). Also, we investigated the possibility of using the intramuscular electrode for reducing tremor in one patient by electrical stimulation delivered with timing based on the electromyographic activity recorded with the same electrode. Main results. In the tremor patients, the current corresponding to the perception threshold and the current limit were 0.7 ± 0.2 and 1.4 ± 0.7 mA for the wrist flexor muscles and 0.4 ± 0.2 and 1.5 ± 0.7 mA for the extensors. In one patient, closed-loop stimulation resulted in a decrease of the tremor power >50%. In healthy individuals the perception threshold and current limits were 0.9 ± 0.6 and 2.1 ± 0.6 mA for the extensor carpi radialis muscle. The subjects could distinguish four or six stimulation patterns (two or three stimulation sites × two stimulation current amplitudes) with true positive rate >80% (two subjects) and >60% (one subject), respectively. Significance. The proposed electrode provides a compact multichannel interface for recording electromyogram and delivering electrical stimulation in applications such as neuroprostheses for tremor suppression and closed-loop myoelectric prostheses.
KW - electrical stimulation
KW - intramuscular electrode
KW - neuroprosthesis
KW - sensory feedback
KW - tremor
UR - http://www.scopus.com/inward/record.url?scp=85062833088&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85062833088&partnerID=8YFLogxK
U2 - 10.1088/1741-2552/ab047a
DO - 10.1088/1741-2552/ab047a
M3 - Article
C2 - 30721892
AN - SCOPUS:85062833088
SN - 1741-2560
VL - 16
JO - Journal of Neural Engineering
JF - Journal of Neural Engineering
IS - 2
M1 - 026035
ER -