Abstract
This study focused on the potential risks of radiofrequency-induced heating of cardiac implantable electronic devices (CIEDs) in children and adults with epicardial and endocardial leads of varying lengths during cardiothoracic MRI scans. Infants and young children are the primary recipients of epicardial CIEDs, though the devices have not been approved as MR conditional by the FDA due to limited data, leading to pediatric hospitals either refusing the MRI service to most pediatric CIED patients or adopting a scan-all strategy based on results from adult studies. The study argues that risk–benefit decisions should be made on an individual basis. We used 120 clinically relevant epicardial and endocardial device configurations in adult and pediatric anthropomorphic phantoms to determine the temperature rise during RF exposure at 1.5 T. The results showed that there was significantly higher RF heating of epicardial leads than endocardial leads in the pediatric phantom, but not in the adult phantom. Additionally, body size and lead length significantly affected RF heating, with RF heating up to 12 °C observed in models based on younger children with short epicardial leads. The study provides evidence-based knowledge on RF-induced heating of CIEDs and highlights the importance of making individual risk–benefit decisions when assessing the potential risks of MRI scans in pediatric CIED patients.
Original language | English (US) |
---|---|
Article number | 2847 |
Journal | Diagnostics |
Volume | 13 |
Issue number | 17 |
DOIs | |
State | Published - Sep 2023 |
Keywords
- MRI
- cardiac
- epicardial leads
- imaging
- implants
- pediatric
- safety
ASJC Scopus subject areas
- Clinical Biochemistry