Comparability of Two Commercially Available Artificial Mastoids

Laura Ann Wilber*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

The new ANSI Standard S3.13-1972 for an artificial headbone lists in its appendix suggested threshold values for audiometric bone conduction calibration using a Beltone artificial mastoid. Two sets of comparisons were recently made between the Beltone 5A and the Brüel & Kjær 4930 artificial mastoids. The average values of these comparisons are given to provide numbers which may be related to those in the ANSI appendix when using a Bröel & Kjaer artificial mastoid.

Original languageEnglish (US)
Pages (from-to)1265-1266
Number of pages2
Journaljournal of the Acoustical Society of America
Volume52
Issue number4B
DOIs
StatePublished - Oct 1972

ASJC Scopus subject areas

  • Arts and Humanities (miscellaneous)
  • Acoustics and Ultrasonics

Fingerprint

Dive into the research topics of 'Comparability of Two Commercially Available Artificial Mastoids'. Together they form a unique fingerprint.

Cite this