Dosimetric verification of gynecological implants

B. Arjomandy*, M. Rajotte, I. J. Das

*Corresponding author for this work

Research output: Contribution to journalConference articlepeer-review

Abstract

A simple analytical method for the accurate calculation of the relationship between bladder and rectum dose with respect to the source strength is presented. The method can be used for a quick check for the computer-generated results as well as it can be used for simple quality assurance method for dose calculation. This method is accurate to within 3%.

Original languageEnglish (US)
Pages (from-to)2771-2773
Number of pages3
JournalAnnual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings
Volume4
StatePublished - 2000
Externally publishedYes
Event22nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Chicago, IL, United States
Duration: Jul 23 2000Jul 28 2000

ASJC Scopus subject areas