N2 - We studied the cause of hemolysis during extracorporeal membrane oxygenation (ECMO) by monitoring hematologic and coagulation profiles in seven consecutive infants treated with this procedure. A constrained vortex pump was used in all patients, and the average duration of ECMO was 224±111 (SD) hours. In all patients, plasma free hemoglobin was low during the first 48 hours after the initiation of ECMO. Later, when visible clots appeared in the ECMO circuit, plasma hemoglobin progressively rose. A rise in the level of fibrin degradation products and a fall in the fibrinogen level were observed concurrently with a rise in the plasma hemoglobin level. After complete circuit changes in six patients, plasma free hemoglobin, fibrin split products, and fibrinogen all returned to baseline values. Neither circuit component changes nor exchange transfusion was effective in normalizing the levels of plasma free hemoglobin, fibrin split products, and fibrinogen. We conclude that when ECMO is administered for prolonged periods, circuit thrombosis occurs and hemolysis ensues. Additional studies are needed to assess the contribution of the constrained vortex pump to this process.
AB - We studied the cause of hemolysis during extracorporeal membrane oxygenation (ECMO) by monitoring hematologic and coagulation profiles in seven consecutive infants treated with this procedure. A constrained vortex pump was used in all patients, and the average duration of ECMO was 224±111 (SD) hours. In all patients, plasma free hemoglobin was low during the first 48 hours after the initiation of ECMO. Later, when visible clots appeared in the ECMO circuit, plasma hemoglobin progressively rose. A rise in the level of fibrin degradation products and a fall in the fibrinogen level were observed concurrently with a rise in the plasma hemoglobin level. After complete circuit changes in six patients, plasma free hemoglobin, fibrin split products, and fibrinogen all returned to baseline values. Neither circuit component changes nor exchange transfusion was effective in normalizing the levels of plasma free hemoglobin, fibrin split products, and fibrinogen. We conclude that when ECMO is administered for prolonged periods, circuit thrombosis occurs and hemolysis ensues. Additional studies are needed to assess the contribution of the constrained vortex pump to this process.
UR - http://www.scopus.com/inward/record.url?scp=0024419491&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0024419491&partnerID=8YFLogxK
U2 - 10.1016/S0022-3476(89)80299-9
DO - 10.1016/S0022-3476(89)80299-9
M3 - Article
C2 - 2795361
AN - SCOPUS:0024419491
SN - 0022-3476
VL - 115
SP - 625
EP - 630
JO - The Journal of pediatrics
JF - The Journal of pediatrics