In order to seal off an experimental cell containing liquid He at temperatures down to 20 mK and at pressures up to 25 atm we have developed a valve that is actuated hydraulically by liquid He and which seals repeatedly with a leak rate of less than 10-10 cm3/sec of liquid He II.
Original language | English (US) |
---|---|
Pages (from-to) | 898-899 |
Number of pages | 2 |
Journal | Review of Scientific Instruments |
Volume | 43 |
Issue number | 6 |
DOIs | |
State | Published - Dec 1 1972 |