Inter-rater agreement of novel high-resolution impedance manometry metrics: Bolus flow time and esophageal impedance integral ratio

D. A. Carlson*, Z. Lin, W. Kou, J. E. Pandolfino

*Corresponding author for this work

Research output: Contribution to journalComment/debatepeer-review

19 Scopus citations

Abstract

Background: Novel high-resolution impedance manometry (HRIM) metrics of bolus flow time (BFT) and esophageal impedance integral (EII) ratio have demonstrated clinical utility, though the reliability of their analysis has not been assessed. We aimed to evaluate the inter-rater agreement of the BFT and EII ratio. Methods: HRIM studies including five upright, liquid swallows from 40 adult patients were analyzed by two raters using a customized MATLAB program to generate the BFT and EII ratio. Inter-rater agreement was assessed using the intraclass correlation coefficient (ICC) for median values generated per patient and also for all 200 swallows. Key Results: The ICC (95% confidence interval, CI) for BFT was 0.873 (0.759-0.933) for median values and 0.838 (0.778-0.881) for all swallows. The ICC (95% CI) for EII ratio was 0.983 (0.968-0.991) for median values and 0.905 (0.875-0.928) for all swallows. Median values for both BFT and EII ratio were similar between the two raters (P-values.05). Conclusions and Inferences: The BFT and EII ratio can be reliably calculated as supported by generally excellent inter-rater agreement. Thus, broader utilization of these measures appears feasible and would facilitate further evaluation of their clinical utility.

Original languageEnglish (US)
Article numbere13289
JournalNeurogastroenterology and Motility
Volume30
Issue number6
DOIs
StatePublished - Jun 2018

Keywords

  • achalasia
  • dysphagia
  • high-resolution manometry
  • impedance
  • reliability

ASJC Scopus subject areas

  • Endocrine and Autonomic Systems
  • Gastroenterology
  • Physiology

Fingerprint

Dive into the research topics of 'Inter-rater agreement of novel high-resolution impedance manometry metrics: Bolus flow time and esophageal impedance integral ratio'. Together they form a unique fingerprint.

Cite this