T1 - Intraoperative measurement of intraocular pressure in vitrectomized aphakic air-filled eyes using the Tono-Pen XL
AU - Badrinath, S. S.
AU - Vasudevan, R.
AU - Murugesan, R.
AU - Basti, S.
AU - Nicholson, A. D.
AU - Singh, P.
AU - Gopal, L.
AU - Sharma, T.
AU - Rao, S. B.
AU - Abraham, C.
PY - 1993/1/1
Y1 - 1993/1/1
N2 - The Tono-Pen XL (Bio-Rad, Santa Ana, CA) was compared with manometer readings for intraoperative measurement of intraocular pressure (IOP) in 40 eyes of 40 consecutive patients after vitrectomy, lensectomy, and fluid-air exchange. Tono-Pen readings corresponding to manometer readings of 10, 20, 30, 40, and 50 mmHg were obtained in a masked fashion with a randomized sequence of manometer readings. A correlation was obtained between the manometer and Tono-Pen readings (r = 0.96 in emmetropic eyes and r = 0.93 in myopic eyes). The regression curve that represents the calibration curve of Tono-Pen in terms of the manometer readings for air-filled vitrectomized eyes was obtained. Any Tono-Pen reading can be easily translated into the corresponding manometer reading by referring to the curve. The Tono-Pen can therefore be effectively used to accurately determine intraoperative IOP in eyes undergoing vitrectomy, lensectomy, and fluid-air exchange.
AB - The Tono-Pen XL (Bio-Rad, Santa Ana, CA) was compared with manometer readings for intraoperative measurement of intraocular pressure (IOP) in 40 eyes of 40 consecutive patients after vitrectomy, lensectomy, and fluid-air exchange. Tono-Pen readings corresponding to manometer readings of 10, 20, 30, 40, and 50 mmHg were obtained in a masked fashion with a randomized sequence of manometer readings. A correlation was obtained between the manometer and Tono-Pen readings (r = 0.96 in emmetropic eyes and r = 0.93 in myopic eyes). The regression curve that represents the calibration curve of Tono-Pen in terms of the manometer readings for air-filled vitrectomized eyes was obtained. Any Tono-Pen reading can be easily translated into the corresponding manometer reading by referring to the curve. The Tono-Pen can therefore be effectively used to accurately determine intraoperative IOP in eyes undergoing vitrectomy, lensectomy, and fluid-air exchange.
UR - http://www.scopus.com/inward/record.url?scp=0027716118&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0027716118&partnerID=8YFLogxK
U2 - 10.1097/00006982-199313040-00007
DO - 10.1097/00006982-199313040-00007
M3 - Article
C2 - 8115731
AN - SCOPUS:0027716118
SN - 0275-004X
VL - 13
SP - 307
EP - 311
JO - Retina
JF - Retina
IS - 4
ER -