TY - JOUR
T1 - Just how stable are escape rhythms after atrioventricular junction ablation?
AU - Arora, Rishi K
AU - Spatz, Erica
AU - Vijayaraman, Pugazhendhi
AU - Rosengarten, Michael
AU - Gross, Jay
AU - Kim, Soo
AU - Fisher, John
AU - Ferrick, Kevin J.
PY - 2010/8/1
Y1 - 2010/8/1
N2 - Background: Atrioventricular (AV) node ablation with implantation of a permanent pacemaker is an established mode of therapy in the treatment of atrial fibrillation. However, concern exists regarding subsequent dependency on an entirely paced rhythm and the possible sequela of unheralded pacemaker failure. Data regarding escape rhythm lability, an important feature of pacemaker dependency, are limited. Aims and Methods: The purpose of this study was twofold: (1) to determine the characteristics of escape rhythms at predefined serial time intervals following AV node ablation and pacemaker implantation, and (2) to identify risk factors predictive of unstable escape rhythms. Patients undergoing AV node ablation and pacemaker implantation were assessed for the presence or absence of an escape rhythm during pacemaker interrogation at five predetermined serial time points. Baseline demographics and comorbid conditions were evaluated as potential predictors of those with labile escape rhythms. Results: Seventy-nine percent of the 96 patients studied had an underlying escape rhythm (≥30 beats per minute) immediately postablation. Although the percentage of patients with an escape rhythm increased at each follow-up interval, the number of patients who consistently demonstrated an escape rhythm declined with each follow-up, with 28% of patients lacking an escape rhythm at some time point, i.e., labile escape rhythm. There were no significant predictors of a labile escape rhythm. Conclusion: Among patients who have undergone AV node ablation and pacemaker implantation, 72% have a stable escape rhythm over time, but others are at risk for pacemaker dependency, as predicted by an underlying absent or labile escape rhythm.
AB - Background: Atrioventricular (AV) node ablation with implantation of a permanent pacemaker is an established mode of therapy in the treatment of atrial fibrillation. However, concern exists regarding subsequent dependency on an entirely paced rhythm and the possible sequela of unheralded pacemaker failure. Data regarding escape rhythm lability, an important feature of pacemaker dependency, are limited. Aims and Methods: The purpose of this study was twofold: (1) to determine the characteristics of escape rhythms at predefined serial time intervals following AV node ablation and pacemaker implantation, and (2) to identify risk factors predictive of unstable escape rhythms. Patients undergoing AV node ablation and pacemaker implantation were assessed for the presence or absence of an escape rhythm during pacemaker interrogation at five predetermined serial time points. Baseline demographics and comorbid conditions were evaluated as potential predictors of those with labile escape rhythms. Results: Seventy-nine percent of the 96 patients studied had an underlying escape rhythm (≥30 beats per minute) immediately postablation. Although the percentage of patients with an escape rhythm increased at each follow-up interval, the number of patients who consistently demonstrated an escape rhythm declined with each follow-up, with 28% of patients lacking an escape rhythm at some time point, i.e., labile escape rhythm. There were no significant predictors of a labile escape rhythm. Conclusion: Among patients who have undergone AV node ablation and pacemaker implantation, 72% have a stable escape rhythm over time, but others are at risk for pacemaker dependency, as predicted by an underlying absent or labile escape rhythm.
KW - atrioventricular node ablation
KW - escape rhythm
KW - pacemaker dependency
UR - http://www.scopus.com/inward/record.url?scp=77955285796&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77955285796&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8159.2010.02756.x
DO - 10.1111/j.1540-8159.2010.02756.x
M3 - Article
C2 - 20487338
AN - SCOPUS:77955285796
SN - 0147-8389
VL - 33
SP - 939
EP - 944
JO - PACE - Pacing and Clinical Electrophysiology
JF - PACE - Pacing and Clinical Electrophysiology
IS - 8
ER -