Abstract
We proposed a new approach to improve the control of prosthetic arm rotation in amputees. Arm rotation is sensed by implanting a small permanent magnet into the distal end of the residual bone, which produces a magnetic field. The position of the bone rotation can be derived from magnetic field distribution detected with magnetic sensors on the arm surface, and then conveyed to the prosthesis controller to manipulate the rotation of the prosthesis. Proprioception remains intact for residual limb skeletal structures; thus, this control system should be natural and easy-to-use. In this study, simulations have been conducted in an upper arm model to assess the feasibility and performance of sensing the voluntary rotation of residual humerus with an implanted magnet. A sensitivity analysis of the magnet size and arm size was presented. The influence of relative position of the magnet to the magnetic sensors, orientation of the magnet relative to the limb axis, and displacement of the magnetic sensors on the magnetic field was evaluated. The performance of shielding external magnetostatic interference was also investigated. The simulation results suggest that the direction and angle of rotation of residual humerus could be obtained by decoding the magnetic field signals with magnetic sensors built into a prosthetic socket. This pilot study provides important guidelines for developing a practical interface between the residual bone rotation and the prosthesis for control of prosthetic rotation.
Original language | English (US) |
---|---|
Article number | 6 |
Pages (from-to) | 2134-2142 |
Number of pages | 9 |
Journal | IEEE Transactions on Biomedical Engineering |
Volume | 55 |
Issue number | 9 |
DOIs | |
State | Published - Sep 2008 |
Funding
Manuscript received July 29, 2007; revised February 28, 2008. This work was supported by the National Institutes of Health (NIH), National Institute of Child and Human Development, under Grant R01 HD043137-01. Asterisk indicates corresponding author.
Keywords
- Computer simulation
- Finite-element (FE) model
- Magnetic field
- Permanent magnet
- Prosthesis
- Upper limb amputation
ASJC Scopus subject areas