New time-of-flight electronics for atom-probe field-ion microscopy

D. K. Chan*, B. M. Davis, David N Seidman

*Corresponding author for this work

Research output: Contribution to journalArticle

5 Citations (Scopus)

Abstract

We present new time-of-flight electronics for atom-probe field-ion microscopy that is based on the LeCroy 2277/EXP2 time-to-digital converter (TDC) and is significantly superior to the commonly used LeCroy 4208 TDC. The maximum number of ions detected for each cycle is increased, the electronics dead time is decreased, and the pulse width from the Phillips Scientific 6904 discriminator is measured. The LeCroy 2277/EXP2 TDC records up to 128 ions per field evaporation pulse as opposed to a maximum of eight ions for a single LeCroy 4208 TDC. The dead time is reduced by increasing the pulse pair resolution, in the multihit mode of the TDC, from 7.3 to 3.3 ns. The LeCroy 2277/EXP2 TDC has both leading and trailing edge detection capability, allowing us to measure the pulse width generated by the Phillips Scientific 6904 discriminator. Combining the pulse width measurement capability with the updating capability of the Phillips Scientific 6904 discriminator, in which the discriminator extends the output pulse width by the time difference between two closely spaced (<2 ns) pulses, we resolve two ions arriving within 2 ns of each other. With these improvements, we find that 66% of the total number of multiply field-evaporated ions arriving with a pulse separation of up to 8 ns are incorrectly counted as a single ion. The percentage of multiply field-evaporated ions occurring with a pulse separation of 2 ns is 6%.

Original languageEnglish (US)
Pages (from-to)1973-1977
Number of pages5
JournalReview of Scientific Instruments
Volume65
Issue number6
DOIs
StatePublished - Dec 1 1994

Fingerprint

Microscopic examination
Electronic equipment
microscopy
converters
Atoms
discriminators
Discriminators
probes
Ions
electronics
atoms
ions
pulse duration
pulses
edge detection
trailing edges
Edge detection
leading edges
Evaporation
evaporation

ASJC Scopus subject areas

  • Instrumentation
  • Physics and Astronomy (miscellaneous)

Cite this

@article{340d22ec63884173860afdf4c4cf4ebd,
title = "New time-of-flight electronics for atom-probe field-ion microscopy",
abstract = "We present new time-of-flight electronics for atom-probe field-ion microscopy that is based on the LeCroy 2277/EXP2 time-to-digital converter (TDC) and is significantly superior to the commonly used LeCroy 4208 TDC. The maximum number of ions detected for each cycle is increased, the electronics dead time is decreased, and the pulse width from the Phillips Scientific 6904 discriminator is measured. The LeCroy 2277/EXP2 TDC records up to 128 ions per field evaporation pulse as opposed to a maximum of eight ions for a single LeCroy 4208 TDC. The dead time is reduced by increasing the pulse pair resolution, in the multihit mode of the TDC, from 7.3 to 3.3 ns. The LeCroy 2277/EXP2 TDC has both leading and trailing edge detection capability, allowing us to measure the pulse width generated by the Phillips Scientific 6904 discriminator. Combining the pulse width measurement capability with the updating capability of the Phillips Scientific 6904 discriminator, in which the discriminator extends the output pulse width by the time difference between two closely spaced (<2 ns) pulses, we resolve two ions arriving within 2 ns of each other. With these improvements, we find that 66{\%} of the total number of multiply field-evaporated ions arriving with a pulse separation of up to 8 ns are incorrectly counted as a single ion. The percentage of multiply field-evaporated ions occurring with a pulse separation of 2 ns is 6{\%}.",
author = "Chan, {D. K.} and Davis, {B. M.} and Seidman, {David N}",
year = "1994",
month = "12",
day = "1",
doi = "10.1063/1.1144798",
language = "English (US)",
volume = "65",
pages = "1973--1977",
journal = "Review of Scientific Instruments",
issn = "0034-6748",
publisher = "American Institute of Physics Publising LLC",
number = "6",

}

New time-of-flight electronics for atom-probe field-ion microscopy. / Chan, D. K.; Davis, B. M.; Seidman, David N.

In: Review of Scientific Instruments, Vol. 65, No. 6, 01.12.1994, p. 1973-1977.

Research output: Contribution to journalArticle

TY - JOUR

T1 - New time-of-flight electronics for atom-probe field-ion microscopy

AU - Chan, D. K.

AU - Davis, B. M.

AU - Seidman, David N

PY - 1994/12/1

Y1 - 1994/12/1

N2 - We present new time-of-flight electronics for atom-probe field-ion microscopy that is based on the LeCroy 2277/EXP2 time-to-digital converter (TDC) and is significantly superior to the commonly used LeCroy 4208 TDC. The maximum number of ions detected for each cycle is increased, the electronics dead time is decreased, and the pulse width from the Phillips Scientific 6904 discriminator is measured. The LeCroy 2277/EXP2 TDC records up to 128 ions per field evaporation pulse as opposed to a maximum of eight ions for a single LeCroy 4208 TDC. The dead time is reduced by increasing the pulse pair resolution, in the multihit mode of the TDC, from 7.3 to 3.3 ns. The LeCroy 2277/EXP2 TDC has both leading and trailing edge detection capability, allowing us to measure the pulse width generated by the Phillips Scientific 6904 discriminator. Combining the pulse width measurement capability with the updating capability of the Phillips Scientific 6904 discriminator, in which the discriminator extends the output pulse width by the time difference between two closely spaced (<2 ns) pulses, we resolve two ions arriving within 2 ns of each other. With these improvements, we find that 66% of the total number of multiply field-evaporated ions arriving with a pulse separation of up to 8 ns are incorrectly counted as a single ion. The percentage of multiply field-evaporated ions occurring with a pulse separation of 2 ns is 6%.

AB - We present new time-of-flight electronics for atom-probe field-ion microscopy that is based on the LeCroy 2277/EXP2 time-to-digital converter (TDC) and is significantly superior to the commonly used LeCroy 4208 TDC. The maximum number of ions detected for each cycle is increased, the electronics dead time is decreased, and the pulse width from the Phillips Scientific 6904 discriminator is measured. The LeCroy 2277/EXP2 TDC records up to 128 ions per field evaporation pulse as opposed to a maximum of eight ions for a single LeCroy 4208 TDC. The dead time is reduced by increasing the pulse pair resolution, in the multihit mode of the TDC, from 7.3 to 3.3 ns. The LeCroy 2277/EXP2 TDC has both leading and trailing edge detection capability, allowing us to measure the pulse width generated by the Phillips Scientific 6904 discriminator. Combining the pulse width measurement capability with the updating capability of the Phillips Scientific 6904 discriminator, in which the discriminator extends the output pulse width by the time difference between two closely spaced (<2 ns) pulses, we resolve two ions arriving within 2 ns of each other. With these improvements, we find that 66% of the total number of multiply field-evaporated ions arriving with a pulse separation of up to 8 ns are incorrectly counted as a single ion. The percentage of multiply field-evaporated ions occurring with a pulse separation of 2 ns is 6%.

UR - http://www.scopus.com/inward/record.url?scp=0005251103&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0005251103&partnerID=8YFLogxK

U2 - 10.1063/1.1144798

DO - 10.1063/1.1144798

M3 - Article

VL - 65

SP - 1973

EP - 1977

JO - Review of Scientific Instruments

JF - Review of Scientific Instruments

SN - 0034-6748

IS - 6

ER -