Abstract
The spectral distributions of the radiation emitted by five visible range curing lights were measured spectroradiometrically. The light intensities in specific wavelength bands were evaluated by graphical integration. The results were then compared with the instantaneous readings of power density of all of the units as measured with a commercial curing radiometer. The data obtained by both techniques correlated well at narrower specific wavelengths (450-500 nm), which are assumed to be more effective for photopolymerization. This finding indicates that curing radiometer is sensitive to the desired wavelengths and may be an effective tool to quickly characterize the curing efficiency of dental photocuring sources in a clinical setting.
Original language | English (US) |
---|---|
Pages (from-to) | 373-377 |
Number of pages | 5 |
Journal | Journal of Dentistry |
Volume | 21 |
Issue number | 6 |
DOIs | |
State | Published - Dec 1993 |
Keywords
- Photopolymerization
- Power density
- Spectral intensity
- Wavelength